2/10-ton IR Automatic Tablets Press Machine SAT-2A/10A

Instrument model:SAT-2A/10A

Pressure limit:0.3-2T/10A(1-28MPa)

Automatic control:Sayta intelligent program is used to controlthe five-stage pressurization

Pressure mode:Automatic pressure replenishment-timedpressure retention -automatic pressure relief

Holding time:Unlimited time (seconds)

Pressure conversion:The system automatically converts the pressure of the sample

Pressure conversion:7 inch touch screen

Equipment protection:100mm high-strength aluminum oxide plate+6mm plexiglass door

Limit protection:Cylinder above the limit height automaticpressure relief

Host protection:The actual pressure exceeds the maximumpressure of the machine and automaticallystarts the pressure relief program

Data administration:Can store 10,000 historical operation records,and can be exported to the U disk

Cylinder path:≤30mm

Work space:145 X160mm

Size:230x450x420mm

Device power supply:220V/60Hz

Weight:85Kg

Product Detail

Reliable Pressure Solution for High-Precision FTIR Spectroscopy Samples

The accuracy of FTIR (Fourier Transform Infrared Spectroscopy) analysis depends heavily on the quality of sample pellets. Inconsistent pressure, uneven compression, or air gaps in pellets can directly affect spectral clarity and lead to unreliable test results.

To solve this challenge, the 2–10 Ton IR Automatic Tablet Press Machine (SAT-2A / SAT-10A) is designed specifically for laboratory-grade sample preparation. It provides stable hydraulic pressure, precise control, and consistent repeatability for KBr pellet formation and infrared spectroscopy testing.

Whether used in chemical research, pharmaceutical development, polymer analysis, or material science laboratories, this automatic tablet press ensures high-quality sample preparation every time.

2-ton IR automatic tablet press

What is an FTIR Tablet Press Machine?

An FTIR tablet press machine is a laboratory hydraulic press used to compress powdered samples into solid pellets for infrared spectroscopy analysis.

In FTIR testing, materials such as potassium bromide (KBr) are mixed with samples and compressed into transparent pellets. These pellets allow infrared light to pass through and generate accurate spectral data.

Compared with manual pressing tools, an automatic FTIR tablet press provides:

  • Stable and adjustable pressure output
  • Higher repeatability for laboratory testing
  • Reduced operator error
  • Improved pellet transparency and uniformity

This makes it an essential instrument in modern analytical laboratories.

Why Sample Quality Matters in FTIR Analysis

Accurate FTIR results depend heavily on the quality of pellet preparation. If pressure is unstable or uneven, it may cause:

  • Weak or distorted absorption peaks
  • Inconsistent sample thickness
  • Reduced infrared transmission efficiency
  • Poor repeatability in experimental data

The SAT-2A / SAT-10A ensures controlled pressure from 2 to 10 tons, enabling stable pellet density and improving analytical accuracy across repeated tests.

Key Features of SAT-2A / SAT-10A Automatic Tablets Press

The machine is designed with laboratory precision and industrial reliability in mind.

  • Stable Hydraulic Pressure System: Provides adjustable pressure from 2 to 10 tons, ensuring accurate and repeatable sample compression.
  • Automatic Pressing Operation: Reduces manual labor and improves efficiency in laboratory workflows.
  • Digital Pressure Control Interface: Allows users to set and monitor pressure levels easily during operation.
  • High Safety Protection System: Includes overload protection, pressure release safety mechanism, and operational limit control.
  • Durable Steel Structure: Built with high-strength materials for long-term laboratory use and stability.
  • Precise Mold Compatibility: Compatible with standard FTIR pellet dies for consistent sample preparation.

Application of 2-ton automatic tablet press

Product Description

Item Specification
Instrument model SAT-2A
Pressure limit 0.3-2T (1-28MPa)
Automatic control Savta intelligent program is used to control the five-stage pressurization
Pressure mode Automatic pressure replenishment - timed pressure retention - automatic pressure relief
Holding time Unlimited time (seconds)
Pressure conversion The system automatically converts the pressure of the sample
Pressure conversion 7 inch touch screen
Equipment protection 100mm high-strength aluminum oxide plate + 6mm plexiglass door
Limit protection Cylinder above the limit height automatic pressure relief
Host protection The actual pressure exceeds the maximum pressure of the machine and automatically starts the pressure relief program
Data administration Can store 10,000 historical operation records, and can be exported to the U disk
Cylinder path ≤30mm
Work space 145 × 160mm
Size 230 × 450 × 420mm
Device power supply 220V/60Hz
Weight 85Kg

Applications of IR Tablet Press Machine

The SAT-2A / SAT-10A is widely used in scientific and industrial laboratories for FTIR sample preparation and material testing applications.

FTIR Spectroscopy Sample Preparation

Used for compressing KBr and sample mixtures into transparent pellets for infrared spectroscopy analysis. It ensures stable infrared transmission and consistent spectral results in research and routine laboratory testing.

Pharmaceutical Laboratory Testing

Applied in pharmaceutical analysis for raw material identification, compound structure evaluation, and formulation consistency testing. It supports reliable FTIR-based quality control processes.

Chemical Research Laboratories

Used for preparing solid samples in organic and inorganic chemical analysis. It supports molecular structure identification, reaction product verification, and spectroscopic research experiments.

Polymer and Material Science

Used in polymer composition analysis, additive detection, and material performance studies. It ensures uniform pellet formation for accurate FTIR evaluation of complex materials.

Academic Research Institutions

Widely used in universities and research institutes for FTIR teaching experiments and scientific research projects. It provides stable sample preparation for spectroscopy training and laboratory demonstrations.

Quality Control Laboratories

Used in industrial QC environments for routine FTIR testing of raw materials and production samples. It ensures consistent sample preparation for reliable batch comparison and quality verification.

How the FTIR Tablet Press Works

The operation process is simple and efficient, suitable for laboratory environments:

  1. Sample Loading
    The powdered sample and KBr mixture are placed into the mold cavity.
  2. Pressure Setting
    The operator selects the required pressure level based on sample requirements.
  3. Automatic Compression
    The hydraulic system applies uniform pressure to form a dense pellet.
  4. Pressure Holding & Release
    Pressure is maintained for a set time to ensure structural stability, then released automatically.
  5. Sample Removal
    The finished pellet is removed and ready for FTIR analysis.

This automated workflow ensures consistent sample quality with minimal operator variation.

Manual vs Automatic Tablet Press

Feature Manual Press SAT-2A / SAT-10A Automatic Press
Pressure Stability Low High
Repeatability Inconsistent Excellent
Operator Effort High Low
Efficiency Low High
Safety Level Medium High
FTIR Result Accuracy Variable Stable

The automatic system significantly improves both laboratory efficiency and analytical reliability.

Introduction to the Automatic Hydraulic Machine Interface 1

Operating Procedure

The SAT-2A / SAT-10A automatic tablet press is designed for simple and stable operation in laboratory environments. The standard operating procedure is as follows:

1. Sample Loading and Mold Assembly

Load the powdered sample (typically mixed with KBr for FTIR applications) into the mold cavity. Assemble the mold components securely to ensure proper sealing and uniform pressure distribution during pressing.

2. Mold Installation in Hydraulic Press

Place the assembled mold into the hydraulic press chamber. Tighten the lead screw to ensure the mold is correctly positioned and fixed before starting the pressing process.

3. Pressure Setting Configuration

Enter the control interface and access the pressure setting screen. Select the required pressure value according to the sample type and testing requirements.

4. Mold Parameter Adjustment

Set the mold specifications according to the sample shape requirements. Adjust the mold diameter settings to ensure compatibility with the FTIR pellet size standard.

5. Start Pressing Operation

Return to the main control interface and press the “Start” button. The system will begin automatic pressure increase according to the preset parameters.

6. Pressure Holding and Automatic Release

When the target pressure is reached, the system will automatically maintain stable pressure for a preset duration to ensure proper pellet formation. After completion, the system will automatically release pressure.

7. Sample Removal

Remove the mold from the press chamber and carefully extract the formed pellet. Ensure the sample remains intact and free from cracks or deformation.

8. FTIR Analysis

Place the prepared pellet into the infrared spectrometer for FTIR analysis. The high-quality, uniform pellet ensures stable infrared transmission and reliable spectral results.

IR Automatic hydraulic machine operation steps

Automatic hydraulic machine operation steps 2

About Hench Technology

Hench Technology Co., Ltd., based in Tianjin, China, specializes in the design and manufacturing of infrared spectroscopy sample preparation equipment. Their product line includes IR hydraulic presses, molds, heated dies, and full accessory sets. As a trusted hydraulic press supplier, Hench provides durable, high-performance solutions for laboratories worldwide.

   2-ton automatic tablet press set

Related Laboratory Equipment

Leave Your Message

×

Leave Your Message